Phantom Loading Arrangement
1,998 plays

More
In this lecture I have discussed the concept of Phantom loading Arrangement in energy meter. We use Phantom loading for calibration of energy meter because we never calibrate the energy meter at full load because at full load we waste a lot of energy in calibration. And we also saw the difference between Phantom loading and direct loading. And at the end we discuss Summary of this course.

Pawan Chandani is teaching live on Unacademy Plus

Pawan Chandani
Faculty of GATE/ESE (Electrical Engineering) , Also a YouTuber and teaches Electrical Eng. subjects on YouTube | Jony Ive Award Winner

U
Unacademy user
sir g ,aapke sare videos dekh lu ,to kitne question match ho sakte hain..
please correct phase angle value at 08:19 minutes
Pawan Chandani
6 months ago
I know that was wrong but I can't change this....after being uploaded the lesson.
Sir, is it possible can the disc moves in anticlockwise direction with positive supply due to some unknown error ? if Yes, which error will be happen ?
sir aapne Flux ka formula is lecture me galat bata diya. Please check f = Q/P =tan-1[ √3(w1-w2)/w1+w2] pf= cos-1(flux)
Pawan Chandani
a year ago
yes, you are right...it's wrong in summery part....But It's correct when I drived it in two wattmeter method part 3....so thanks for noticing this correction. It's good that you watch the lectures carefully.
sir D arsonaval galvanometer per ek video bna digiye na please
Pawan Chandani
a year ago
I've uploaded the preparation strategy for measurement course where I mentioned the syllabus of measurement for Electrical Engineering.... so remaining part of this course will be discussed accordingly.
sir would you cover measurement part of ECE?
  1. Phantom Loading By- Pawan Chandani


  2. Phantom Loadin Eneray meters are not calibrated at full Loa . The actual 0adin9 arrangement will aste a Jlot of e0ner. Oirect loading Arrangement Energy meters are not calibrated at full load v (220)- Rc= o. IA Pc R current Icc-3A 440o 1980 +11 = IgglW' Pressure Coil 220V t10 hr Direct Loading Arrangement


  3. Phantom doading Atrange ment 9A ate 9A ( )L = 1110 2 20 Current Coil Pressure Coil 220V Phantom Loading Arrangement


  4. Summary of this Course Blondel's Theorem Two Klattmetes Method 1 2.


  5. Eriol due to Mc-Connection Er 01 due to LC-Connection 100% Error due to PC Inductance 1.cosB cos =


  6. Construction k Working of Energy, Meter T, VI sin (A- ) Ideally 90 Friction Compensation -Phatom Loadin Creeping Ertor


  7. Pawan Chandani Ask your Doubt....??? 77? comment RATE REVIEW RECOMMEND & Pawan Chandani To get notifications.. Enroll 49 Add to list